24H全国咨询热线 400-668-3536

导航

/ ABOTU US / 关于韦弗斯

百年企业,百年品牌

  • SF6 Gas Tester (Winfoss-P17)
  • SF6 Gas Tester (Winfoss-P17)

SF6 Gas Tester (Winfoss-P17)

SF6 comprehensive tester

Supported platform: Windows/Linux/Unlix

Software size: 10M

Brand: WINFOSS

Model: Winfoss-P17

Measuring range: Dew point -80 ~ +20℃.

Dew point accuracy: ±0.4 ° C (when the dew point temperature is below 0 ° C, the sensor output is frost point).

Resolution: Dew point 0.1 ° C or 0.1PPM

Accuracy: ±1ppm

Response time: 63% [90% + 20-20 ℃ TDS 5s[45s]

产品详情/ Product details

SF6 comprehensive tester (Winfoss-P17) is a set of humidity, percentage concentration, main decomposition products measurement purity, dew point, H2S, SO2, CO and HF multi-functional analysis instrument, the whole machine adopts the best foreign sensors, humidity using Finland Vaisala sensor; Purity using German HNP sensor; The decomposition products are imported from the United States Dekang SF6 special sensor. The instrument uses a large LCD Chinese display, real-time display of various parameters, the whole fool operation, mass information storage, built-in rechargeable lithium battery, AC and DC dual use. It is rated by the industry as a high-tech, energy-saving and environmentally friendly instrument.

 

SF6 gasComprehensive tester(Winfoss-P17) 

 

    

Product overview

SF6 comprehensive tester (Winfoss-P17C) is a comprehensive detection and analysis instrument that sets humidity, percentage concentration, main decomposition products measurement, etc., by detecting the main products of insulation material cracking inside SF6 electrical equipment: SO2, H2S, CO, CF4 content, quickly and accurately determine SF6 circuit breakers, transformers, GIS and transformers and other electrical equipment internal early failure; At the same time, it can also detect dew point and purity, providing a more sufficient basis for equipment diagnosis. The instrument can be used for decomposition, purity, moisture detection alone, and can also be used for various combination detection. Complete a number of functional tests at a time, greatly saving the gas in the equipment, effectively protecting the environment, reducing pollution, while reducing the user's workload, improve work efficiency. The whole machine adopts foreign superior performance sensor, humidity adopts Finnish Vaisala sensor; Purity using German HNP sensor; The decomposition products are imported from the United States Dekang SF6 special sensor. The instrument uses a large LCD Chinese display, real-time display of various parameters, the whole fool type operation, mass information storage, built-in rechargeable battery, AC and DC dual use. It is rated as an innovative, environmentally friendly instrument by the industry.


1729574660.jpg


1729574674.jpg

 

Functional characteristics

l Graphic menu, color screen display;

l Support data communication; Data management system software;

l With self-test and open calibration function and self-diagnosis function;

l Multi-component detection, high precision, good stability, can simultaneously detect moisture, purity, decomposition (SO2, H2S, CO, and CF4), without interference;

l Automatic flow control, efficient and accurate, effectively reduce SF6 gas emissions;

l Built-in electric air pump, quick flush gas system, reset fast, effective protection sensor;

l With electrochemical sensor zero automatic correction function; The anti-cross interference design improves the accuracy of mixed gas detection.

l Full dew point curve tracking, showing the dew point and humidity under the current state, and automatically converted to 20℃, standard atmospheric pressure humidity;

l Electronic flow display and power display, flow automatic adjustment;

lMeasurement data storage, printing, support file system;

l Dual purpose, built-in lithium battery;

l Simple operation and easy maintenance;

l Small size, light weight, easy to carry.

 

Technical parameter


Parameter number entry

Parameter value

dew-point

Measuring range

Dew point -80 ~ +20℃ (PPMV supported)

Dew point accuracy

± 1℃ (resistance-capacitance method, within a certain range)

(When the dew point temperature is below 0 ° C, the output is frost point)

Response time

63% [90% +20- -20℃ T5S [45s]

Resolution

Dew point 0.1℃ or 0.1µ L/L

repeatability

± At 1℃, the flow rate SF6 is adjusted at 0.5 ~ 0.9L/min

purity

Measuring range

0% ~ 100%

Measurement accuracy

≤ 0.03%

Measuring time

<2 minutes

H2S 

Measuring range

0 ~ 100µ L/L

Measurement accuracy

Measured value ≤ 10µ L/L, error ≤ ± 0.5 & micro; L/L

Measured value > 10µ L/L, error ≤ ± 5%

Measuring time

<2 minutes

Minimum detection amount

≤ 0.1 & micro; L/L

repeatability

Measured value ≤ 10µ L/L,≤ 0.2 & micro; L/L

Measured value > 10µ L/L,≤ 2%

SO2

Measuring range

0 ~ 100µ L/L, 0 ~ 2000µ L/L

Measurement accuracy

Measured value ≤ 10µ L/L, error ≤ ± 0.5 & micro; L/L

Measured value > 10µ L/L, error ≤ ± 5%

Measuring time

<2 minutes

Minimum detection amount

≤ 0.1 & micro; L/L

repeatability

Measured value ≤ 10µ L/L,≤ 0.2 & micro; L/L

Measured value > 10µ L/L,≤ 2%

CO

Measuring range

0 ~ 500µ L/L, 0 ~ 1000µ L/L

Measurement accuracy

± 0.5% Measurement value ≤ 50µ L/L, error ≤ 2.0 & micro; L/L

Measured value > 50µ L/L, error ≤ ± 4%

Measuring time

<2 minutes

Minimum detection amount

≤ 0.1 & micro; L/L

repeatability

≤ 1.5% (≤ 50µ L/L; ≤ 3% (> 50µ L/L)

CF4

Measuring range

0 ~ 100µ L/L

Measurement accuracy

Measured value ≤ 10µ L/L, error ≤ ± 0.5 & micro; L/L

Measured value > 10µ L/L, error ≤ ± 5%

Measuring time

<2 minutes

Minimum detection amount

≤ 0.1 & micro; L/L

repeatability

Measured value ≤ 10µ L/L,≤ 0.2 & micro; L/L

Measured value > 10µ L/L,≤ 2%

Zero drift

When the temperature changes from -20℃ to 50℃, the maximum drift is SO2, H2S, HF≤ ± 0.1 & mu; L/L, CO≤ ± 0.5 & mu; L/L

Display mode

Color LCD display

Operating voltage

220VAC± 10% 50Hz, AC and DC dual use, built-in lithium battery, with display power and overcharge protection, undervoltage warning and protection functions.

Working pressure

0 to 1.0MPa

Output interface

USB1.1 specification

Operating environment

Temperature: -20 ~ +60℃, humidity: 10 ~ 95%RH, no condensation at 20℃

Calibration period

Once a year

Product quality

About 10.5KG

 

Product configuration

l SF6 comprehensive tester 1 set;

l Test pipe 1;

l 1 power cord;

l aluminum alloy shock-proof box 1;

l 1 set of switchgear adapter;

l Chinese manual 1;

l Factory test report and other user data 1 set.

 

Service guarantee

3.png

l Provide on-site training guidance and carry out on-site demonstration operation.

l Provide detailed instructions and user guidance in Chinese.

l One year free warranty, lifelong spare parts support and maintenance.

l User file information management, customer service department every semi-annual return visit and tracking service.

l Hotline 365 days to answer the user's consultation.


最近浏览 / Recently Viewed